Jump to:

Philip Morris

Mineral Fiber Identification Using the Analytical Transmission Electron Microscope

Date: Nov 1978
Length: 21 pages
2063105044-2063105064
Jump To Images
spider_pm 2063105044_5064

Fields

Author
Beaman, D.R.
Walker, H.J.
Type
SCRT, REPORT, SCIENTIFIC
ABST, ABSTRACT
BIBL, BIBLIOGRAPHY
CHAR, CHART, GRAPH, TABLE, MAPS
Area
SOLANA,RICHARD/CENTRAL FILES
Litigation
Fali/Produced
Characteristic
EXTR, EXTRA
Site
R545
Named Organization
Dow Chemical of Canada
Imb
Jaffe Wick
Millipore
Natl Bureau of Standards
Union Carbide
Workshop on Asbestos
Dow Chemical
Author (Organization)
Dow Chemical
Named Person
Ampian
Beaman, D.R.
Beers
Biles
Bradford, E.B.
Cliff
Duncumb
Emerson
File
Galle
Geiss, R.H.
Goldstein
Heinrich, K.
Joy
Knox, W.A.
Kyser, D.F.
Lorimer
Maher
Mcgrath, P.
Mizuhira
Morgan
Parmentier, C.
Peterson, D.J.
Philibert
Reed
Ross
Russ
Samudra
Short
Sprys
Sturkey, L.
Suzuki
Tixier
Wards
Ware
Master ID
2063104795/5283
Related Documents:
Date Loaded
20 Sep 1999
UCSF Legacy ID
bbp52d00

Document Images

Text Control

Highlight Text:

OCR Text Alignment:

Image Control

Image Rotation:

Image Size:

Page 1: bbp52d00 Log in for more options!
National Bureau of Standards Special Publication 506. Proceedings of the Workshop on Asbestos: Definitions and Measurement Methods held at NBS, Gaithersburg, MD, July 18-20, 1977. (Issued November 1978) MINERAL FIBER IDENTIFICATION USING THE ANALYTICAL TRANSMISSION ELECTRON MICROSCOPE D. R. Beaman and H. J. Walker The Dow Chemical Company Midland, Michigan 48640 Abstract In a transmission electron microscope equipped with an energy dispersive spectrometer (EDS), It is possible to obtain the high resolu- tion morphology, crystal structure, and elemental composition of sub- micron mineral fibers, particulate, and thin films. The reliability of fiber analysis is enhanced when fiber identification is based on the nearly simultaneous determination of these three characteristics because each of the individual modes can yield ambiguous information. Energy dispersive spectrometer data can be converted to elemental fiber compositions using known standard spectra or relative sensitivity factors which can be calculated or experimentally determined for a given , instrumental configuration. Calculated and experimental sensitivity : factors are found to agree within 15 percent for photon energies above . 1.5 keV. The relative error in composition calculated from EDS spectra ; will generally be better than 10 percent, but only if the TEM column and ' components have been properly modified to reduce the effects of extraneous x-ray generation and electron scattering. The sources of these problems are described and a procedure for minimizing the effects outlined. Proper aperturing, collimation, selection of materials of construction, and operating conditions can provide useful mineral spectra. It is often necessary to correct for x-ray absorption even in fine mineral fibers, and this may be done using reference standards or sensitivity factors corrected for absorption. The effect of absorption increases rapidly as the difference between the mass-absorption coefficients of the elemental constituents of the mineral increases. Carbon contamination which degrades both EDS spectra and electron diffraction patterns can be minimized by using low current density and short analysis times. Less than 15 percent of the chrysotile fibrils in a standard provided positive selected area electron diffraction patterns (SAED), but up to 50 percent did have the correct layer line spacing. The fraction of fibers providing good diffraction increases rapidly as the number of fibrils in a fiber increases. The reported differences in SAED quality arise primarily because investigators use differing criterion for defining a positive SAED pattern and the fiber size distribution examined varies. Sample preparation methods were reviewed and it was found that condensation washing is only reliable if loss corrections are applied, particularly in the case of amphibole fibers. In spite of the many problems, inter-laboratory end multiple sample reproducibility in the measurement of fiber concentrations can be t30 percent when using good procedures. Key Words: Carbon contamination; electron diffraction; mineral fibers; transmission electron microscope; x-ray spectroscopy. 249
Page 2: bbp52d00 Log in for more options!
Introduction The need to identify and determine the concentration of small mineral fibers in environmental samples provided motivation for the development of the analytical trans- mission electron microscope (ATEM) which consists of a conventional transmission electron microscope (CTEM) equipped with energy dispersive spectroscopy (EDS) and possibly scanning transmission electron microscopy (STEM) capabilities. In such an instrument it is possible to obtain from very small volumes of material high resolution morphology in the TEM or STEM mode, elemental data using the EDS, and structural information for crystalline materials in the selected area electron diffraction (SAED) mode. When identification is based on the nearly simultaneous determination of three quantities-morphology, elemental composition, and crystal structure-the reliability of the analysis is significantly improved because the individual modes sometimes yield ambiguous information. The limita- tions of each mode have been discussed previously [1,2]1. All modes are adversely affected by the presence of adjacent non-fibrous debris and overlaying films. Fibers that are too thin or too thick do not provide sufficiently good SAED patterns for positive identification by comparison with standards. Less than 15 percent of the chrysotile fibrils in a particular standard gave positive SAED patterns. Chrysotile diffraction is further degraded by electron beam bombardment and instrumental contamination. Energy dispersive spectrometry is not a panacea because there are different minerals with similar compositions and elemental substitution is common. Morphology is often compromised by the environment and interfering solids. The hollow-core or tubular appearance of chrysotile is distinctive but often absent and degraded during analysis. It is difficult to establish a protocol for basing identification on three criteria, but when this is done the quality of the analysis is significantly improved. This paper describes some of the difficulties associated with fiber counting in the ATEM with the goal of circumventing the problems. The data from an energy dispersive spectrometer can be converted to chemical concentrations but there is a need to calibrate the instrument and correct for x-ray absorption even in very fine fibers. There are instrumental limitations which degrade EDS spectra but can, to some extent, be avoided. Contamination seriously affects both the EDS spectra and SAED patterns, but there is little that can be done to avoid it in existing instruments other than to understand the problem. The reasons for the controversy concerning the quality of SAED patterns from mineral fibers are examined and criteria suggested for classifying chrysotile SAED patterns. Sample preparation methods are reviewed and some results of inter-laboratory reproducibility are presented. Sample Preparation The three methods of water sample preparation that are commonly used are summarized in table 1 and references 1-6. Water is vacuum filtered through 0.22 pm Millipore or 0.1 pm Nuclepore filters. Nuclepore has the advantage of being smooth and therefore not generating a replicated structure when carbon coated; it has the disadvantages of being prone to fiber loss during handling and sporadic occurrences of non-uniform solids deposition during filtration. Millipore retains fibers well but generates a structured background if carbon coated prior to destruction of the filter structure. In the method of condensation washing [1,2,6], TEM grids with carbon-coated Formvar films are positioned on the Ni support screen of the cold finger in a condensation washer. A piece of Whatman filter paper placed between the TEM grid and the Ni support screen has been shown to reduce fiber loss during solvent extraction [7]. The grids are preconditioned by the application of a few drops of acetone beneath the Ni support screen to prevent warping of the filter section. The filter sections are placed, sample side down, on the TEM grid immediately following pre-conditioning. The Millipore is removed in 10-50 minutes of acetone vapor extraction. The complete procedure and sources of errors are described elsewhere [1,2]. 1Figures in brackets indicate the literature references at the end of this paper. 250 E F.n
Page 3: bbp52d00 Log in for more options!
Table 1. Method of preparing liquids for ATEM analysis. Method + reference filter medium pre-treatment fiber fixation by vacuum evaporation of carbon Jaffe-fusion 3,4 0.22 71m Millipore fused in acetone vapor for 5-10 minutes yes pre-conditioning none extraction filter section on configuration grid on polyurethane in enclosed petri dish Jaffe-wick 5,6 0.1 um Nuclepore none yes 10 uL droplet of solvent onto sample positioned on grid filter section on grid on wire mesh on several layers of filter paper in enclosed petri dish Condensation washing 1,2,6 0.22 pm Millipore none no acetone wetting of grid without filter filter section on grid on cold finger in reflux column solvent acetone chloroform acetone duration of 12 hours 10-24 hours 10-50 minutes extraction In the Jaffe-wick method [5,6], the Nuclepore filter is carbon coated after filtration to fix the solids in place prior to filter extraction. The TEM grid is positioned on a wire mesh placed on several layers of filter paper in a,:tri dish. The carbon coated filter section is positioned on a grid and a 10 p1 droplet of chloroform is added to prevent warping. The layers of filter paper are,saturated with chloroform and the Nuclepore extracted slowly (10-24 hours) in the covered petri dish. In the Jaffe-fusion method [3,4], a portion of the Millipore filter is attached to a glass slide and placed for 5-10 minutes in acetone vapor. This short pre-treatment in acetone destroys the structure of the Millipore and therein avoids the formation of a replicated network structure during carbon coating which would interfere with fiber counting. The fused Millipore on glass is carbon coated and then extracted using acetone in the same manner as in the case of the Jaffe-wick method. One of the prime sources of error in the analysis is the fiber loss which occurs during sample preparation. Condensation washing is a popular method of preparation, but it introduces variability in the results and yields higher fiber losses than Jaffe-type methods [1]. While some investigators have obtained good results with condensation washing [8,9], there are a sufficient number of technique problems [1,2] so that serious differences occur in inter-laboratory comparisons. It is possible to correct for the losses associated with condensation washing using partially-extracted Jaffe samples to determine the total fiber concentration [1]. This requires additional preparation time and TEM analysis. Fortunately the chrysotile losses associated with condensation washing are usually below 20 percent [1] and can be considered insignificant if the duration of wash is less than an hour in a properly controlled washer. We have obtained reproducible results using Jaffe extraction of carbon-coated Nuclepore [2] and loss corrections in conjunction with condensation washing. 251
Page 4: bbp52d00 Log in for more options!
All of the above discussion refers to water samples. In preparing air samples it is preferable to low-temperature ash the filter because of the heavy filter loading associated with air sampling. The ash is then suspended in water and processed as a water sample. Because the ash tends to be clumped, it is necessary to subject the suspended ash to ultrasonic treatment. Instrumental Limitations Instrumental problems arise when using energy dispersive spectrometers, because TEMs were never intended to be used in quantitative chemical analysis and ATEMs have been constructed by retrofitting EDS and STEM capabilities to existing systems. There are two prime sources of the instrumental problem: 1) the EDS is not a focusing spectrometer and is Insensitive to the location of the x-ray source and, thus, will detect all x-rays with a line-of-sight path to the detector [3]; 2) in a typical CTEM column there is, in a confined volume, a high density of hardware such as pole pieces, apertures, anti-contamination surfaces, sample grids, samples holders and associated clips. These two features combine to yield remote x-ray generation, i.e., x-radiation originating from regions outside of the volume excited by the primary electron beam. This causes: 1) spectral peaks unrelated to the sample to appear in the EDS spectrum leading to quantitative inaccuracy and errors in identification; 2) increased background radiation which raises the detectability limits; and 3) a loss in spatial resolution. The sources of the problem are secondary fluorescence by characteristic and continuous radiation generated in the column apertures, backscattered electrons from the sample and its support, and scattered primary electrons. The use of high voltages to penetrate thin samples and retain good spatial resolution leads to the generation of characteristic and continuous radiation in column apertures. The second condenser (C2) variable aperture, which is the last aperture above the sample, poses the most serious problem. The maximum in the generated continuum at a bevrt energy of 100 keV and PtKa characteristic radiation both have wavelengths of abou~ 0.2A and are readily transmitted by thin Pt apertures, e.g., over 40 percent of the 0.2 Pt radiation is transmitted by an 100 pm thick Pt aperture. Most of this radiation will be dissipated by absorption in the column but any that does reach the sample area can generate secondary fluorescence at and near the sample which is unrelated to primary electron beam excitation. Because almost all primary electrons are transmitted by thin films and small particles, the backscattered electron fraction is small as indicated for Au films in figure 1 [11]. If the beam voltage is high and the sample thin, less than 5 percent of the incident electrons will be backscattered. Any electrons that are backscattered toward the detector can penetrate the 7.5 pm Be window of the EDS because they will, for the most part, have energies close to the incident beam energy. Eighty percent of the 100 keV electrons can penetrate 7.5 pm of Be and in so doing lose less than 5 percent of their energy. Most backscattered electrons do not reach the detector because they are confined by the strong objective lens field. They can, however, excite remote particulate matter and the support grid. 252
Page 5: bbp52d00 Log in for more options!
Acceleration Potential In keV Figure 1. The percentage of backscattered electrons as a function pf incident electron energy for two different thicknesses of Au. The data are from Philibert and Tixier [11]. Scattered electrons in the column cause electron beam tailing [12] which leads to excitation of areas in the sample immediately adjacent to the region of primary beam excitation. This effect is due to improper alignment and scattering by column components and increases in severity as the beam voltage is lowered. The following list indicates some steps that may be taken to alleviate these instrumental problems. The magnitude of the problem and, therefore, the effectiveness of these alterations will vary appreciably from one instrument to another because of differences in electron optical configurations, alignment procedures, column cleanliness, aperturing (sizes, materials, thicknesses, and location), and operating mode (TEM vs. STEM). I. Reduce the generation in and transmission of radiation by column apertures. a) Use thick apertures [13] b) Use Pt apertures rather than Mo or Ta [12,14] c) Use column inserts somewhere between C2 and the sample [15] d) The use of low acceleration potential reduces this problem, but promotes beam tailing, backscattering, and absorption effects e) Determine if performance depends upon the emission current for the instrument being used and the type of sample being studied II. Reduce the excitation of material remote to the sample. a) Specimen holders, specimens clamps, and support grids should be made of low atomic number materials (Be, graphite, or polymer) or coated with such materials [1,13,16] b) Use support grids with maximum open area [13] 253 2063105048
Page 6: bbp52d00 Log in for more options!
c) Coat components near the specimen such as anticontamination devices and sample support rods with low atomic number materials (Aquadage) d) The objective aperture must be removed during EDS data acquisition e) The sample support film should be as thin and have as low an atomic number as possible f) Operate at as low a tilt angle as will provide adequate EDS intensities (less area of grid exposed to excitation) III. Optimize the EDS detector configuration. a) Use the greatest Si(Li) crystal-to-sample distance that will provide adequate count rates [17] b) Collimate the detector with a low atomic number material c) The collimator should be thick enough or shielded with sufficient material (high z) to absorb any stray radiation [18] IV. Minimize electron scattering a) b) c) Use a smail (100 pm) condenser aperture [14] Operate at high acceleration potential Have the column clean and properly aligned These effects of extraneous radiation can best be examined by comparing spectra obtained on and off the edge of a thin film or fiber or by comparing the spectra obtained with the beam positioned in a hole (hole-count) [12] with spectra obtained on the sample. In performing on- and off-film measurements on a Sn-Cu-Cr film, 3 percent of the Cr intensity was attributable to Cr plating on the sample hold-down clip while the Cu TEM grid was responsible for 15 percent of the Cu signal. Insertion of an aperture just beneath the variable C2 aperture on a Philips EM300 operated in the TEM mode increased the Cu peak-to-background ratio and reduced the off-film Cu by 35 percent. The maximum peak- to-background ratios have been achieved using a column insert (1 mm ID x 2.57 ms OD x 3mm thick) in the lower end of the vacuum tube through which the variable C2 aperture passes. Kyser and Geiss [18] have found that operation in the STEM mode reduces the extraneous background by about a factor of two. Even after these precautions have been taken, it is still advisable to subtract the off-fiber spectrum from the fiber spectrum and to use as dilute a sample as feasible. A high density of solids on the grid may reduce the analysis time required to find fibers, but it seriously degrades the quality of SAEO patterns and.EDS spectra. Quantitative Analysis There are two aspects to quantitative fiber analysis of environmental samples in the ATEM, namely, the proper identification of the fibers coupled with the accurate determina- tion of the number of fibers per unit area. When the concentration of a specific mineral is sought the best procedure is to compare unknown spectra and diffraction patterns with those obtained from well-characterized standards in the same instrument using constant operating conditions. When unknown samples are encountered, it is advisable to compare ATEM data with the results of x-ray diffraction, infrared spectroscopy, and x-ray fluores- cence in conjunction with a careful consideration of the mineralogy of the problem. When the fibers, particles, or films of interest are thin, the following expression, originally proposed by Duncumb [19] and pursued by Cliff and Lorimer [20] and Russ [21], can provide good results; 254
Page 7: bbp52d00 Log in for more options!
CA IA _ (p-B)A ZB SAS r SAS T"F B (1) where I is the net peak intensity corrected for background and peak overlap and SAB is a relative sensitivity factor, i.e., the ratio of the detected intensities (IB /IA ) for two pure thin standards of the same mass thickness. Absorption, secondary fluorescence, and backscattering effects must be negligible for eq. (1) to be applicable. SAB is most easily measured on multi-element thin standards of known composition. There are not many experimental data and the bulk of what is available has been pub- lished by Cliff and Lorimer [20] and Sprys and Short [22]. SAB can be calculated from the following expression which is fully discussed elsewhere [21,23,24]: SAS 4 AA C10 +ZT) GB 1 n CE~ / EC,A exp' o~ 8e 13.9x10 4, AB C10-~ GA 1 n CE~ ) EC,B exp `- p I Be 13.9x10 4 (2) The subscripts A and B refer to the elements A and B. A is the atomic weight, z is the atomic number, G is the fractional emission in the line of interest, e.g., G(Ka12) °Ko.2 intensity/(Ka12 intensity + KS intensity), Eo is the acceleration energy in keV, Ec ie the excitation energy in keV, and p/plBe is the mass absorption coefficient for A or B radiation by the 7.5 pm Be window on the EDS detector. Note that this expression shows no dependence on the instrumental configuration. However, SAS values determined in different instruments may differ from each other and from theoretical values because: 1) the contribution of secondary fluorescence, back- scattering, and beam tailing may be vastly different in different instruments; 2) the Be window thickness and detector efficiencies may be different and, in some instances, the Si dead layer and Si crystal thickness may be significant; and 3) the samples used to measure SAB may not be truly thin with respect to absorption. Figure 2 compares the values calculated from eq. (2) obtained using the Reed and Ware [25] values for G with the experimental values of Cliff and Lorimer [20]; the ratios are relative to Si, i.e., B = Si. As noted by Goldstein et al. [23] the agreement is poor below 2 keV and good above 2 keV. Table 2 also compares calculated and experimental SAB values. For SMg Si' SA1 Si' STi Si' and SF.e Si, the agreement in the experimental values is generally better than 13 percent (fractional standard deviation or coefficient of variation), notwithstanding the variation in experimental configuration and conditions. With the exception of the SNa Si and SMg Si, the agreement between theory and experiment is better than 15 percent. The SMg Si value determined from eight different mineral fiber standards using the data of Beaman and File [1] was 1.7 ± 0.2 (± 14 percent). This varia- tion is primarily due to inaccuracies in the bulk chemical analysis of the mineral fibers. If IC = 1 and the S values are all relative to Si, n SA,Si1A'i~A Si,Si1i ' 255 (3)
Page 8: bbp52d00 Log in for more options!
Table 2. Calculated and experimental values of the relative sensitivity factor, SA-Si for Ka radiation. Investigator and - - - - - - - - Experimental SA-Si Values--------- Conditions 5Ma-Si SMg-Si SA1-Si STi-Si SFe-Si SCu-Si Cliff & Lorimer[13] EMMA-4 100 kY 5.77 2.07 1.42 1.08 1.27 1.58 0=0° T=45° amphibole particles Beaman & Fiie[2] EM300 80 kV 1.7 ± 0.2 1.4 ± 0.2 1.25 0-39° 7=26° asbestos fibers=0.1 um Sprys & Short[41] EM300 100 kV 7.22 1.08 1.30 silicide particles Morgan et al.[30] EM300 80 kV 3.92 1.55 1.16 1.13 1.38 f<42° 3 ym iso-atomic drops Suzuki et al.[42] JEOL 100C 400 kY 1.7 1.3 2.5 0-0° mineral fibers ---------Calcu7atedSA-SiValues--------- Goldstein et al.[22] 100 kY. 1.66 1.25 1.12 1.16 1.33 1.59 This report Eq.[11] 100 kY 1.52 1.13 1.09 1.07 1.22 1.46 Russ[4] 700 kY 2.01 1.39 1.12 0.95 1.12 1.34 0= tilt angle T - x-ray take-off angle 256
Page 9: bbp52d00 Log in for more options!
Figure 2. Relative sensitivity factors, SA Si, for Ka radiation as a function of the atomic number of element A. The curves are calculated fr.om eq. (2) and the points are experimental values from Cliff and Lorimer [20]; from Beaman [24,i. Other relative sensitivity factors can be calculated from the Si values because SA8/SC8 = SAC. If the 5 values are not relative to Si n CA ° IA/(IA + 1. 5i AIi) . (4) iB ~ e We measured the composition of a 3000A thick Cu-Sn-Cr film on a Cu TEM grid using Philips EM300 CTEM at 80 keV and a Cameca electron probe operated at 25 keV. The results are shown in Table 3 and compared with bulk chemical results. The ATEM results are seriously degraded by the secondary fluorescence and electron scattering as evidenced by the high Cu value resulting from the use of a Cu TEM grid. Off-film spectra were subtracted from the film measurements. The Cr/Sn ratio which is independent of the scattering problems is in good agreement with the chemical data (relative error = 11 percent). The Cu grid was used to demonstrate the difficulties associated with quantitation in the ATEM. As indicated previously, the results will be improved by using low atomic number grids and grids that do not contain any of the elements present in the sample. The results obtained in the electron probe, where scattering problems are minimized by the instrumental configura- tion and the use of low acceleration potential, are excellent (relative error <10 percent). From these limited data and other reported results on thin films [20,26], we conclude that the thin film model of eq. (1) is valid and capable of providing relative errors of less than 10 percent when using experimentally determined 5., values. This represents reasonably good performance when compared with the 5 percent relative error obtained using EDS systems and bulk samples [27]. However, it rust be stressed that this will only be attained in CTEMs after taking the precautions described previously. The accuracy will be best when measuring concentration ratios. The presence of oxide films or organic contamina- tion on the surface and the tendency for surface segregation and particle inhomogeneity to occur complicates and degrades quantitative results. 257 2063105052
Page 10: bbp52d00 Log in for more options!
C7 a Table 3. Experimental composition of a 3000 A thick Cu-Sn-Cr film. Method Element Neutron activation ATEM at 80 keV with SAB values Electron probe at 25 keV with SAB values and absorption corrected Electron probe at 25 keV with 5AB values but no absorption correction Composition in weight percent ~ Cu Sn Cr Cr/Sn 14.6 77.6 7.8 0.101 27 67 6 0.090 15.6 76.7 7.6 0.099 16.4 76.3 7.3 0.096 Correction of Quantitative Data It has generally been assumed that if the sample was transparent to electrons, i.e., structure was visible in the TEM image, then the sample was sufficiently thin so that the only consideration necessary in quantitative analysis was the variation in x-ray generation by the primary electron beam. The loss of ionization through backscattering will generally be negligible for sub-micro diameter mineral fibers, if the acceleration potential is above 80 keV. From figure 1, it is seen that for an 1000A film of Au the voltage could be as low as 50 keV and the backscatter fraction still below 10 percent, whereas over 50 percent would be backscattered by a bulk material. Philibert and Tixier [11] have found that continuous fluorescence is negligible and that characteristic fluorescence will be negligible if p/p ' B line t«l. p/p is the mass Ialloy absorption coefficient for the exciting radiation, B, by the material. It is not presently clear how significant the characteristic fluorescence correction is for thin films because the limited accuracy of the analysis in most CTEMs obscures the effect of characteristic fluorescence. In order to make any corrections to the data, it is necessary to know the thickness which certainly complicates the analysis and detracts from the simplicity of standardless correction. However, for particles and fibers the thickness can often be accurately estimated from the TEM image. Absorption effects in the analysis of mineral fibers were reported by Beaman and File [1] and figure 3 shows the dependence of Ix/ISi on fiber size for various minerals. The ratio of intensity ratios at one fiber radius (rl) to those at another fiber radius (r2) can be determined from Beers law. 258
Page 11: bbp52d00 Log in for more options!
1.0 0.9 0.8 0.7 0.6 0.5 0.4 • • • • • ChrYSOtile Mg/Si Ferroactinolite Ce/Si F• 0.1 Grunerite M Si • n 0.3 0.2 0.1 Ok 0.9 • 0'8 Amosite Fe/Si 0.7 • • • • • 0.3 0.6 • x Hombiende AI/Si 02 •• 0.5 0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.B~ Figure 3. Elemental intensities ratioed to the Si intensity as a function of mineral fiber diameter. The scales for chrysotile, grunerite, and amosite are on the left and on the right for ferroactinolite and hornblende. / x \ 0 exp ' plm pm rl csc y I Ix \,J Ix \, \ S /i \ ~/ rl y C~/ rl eXp (- Plm pm r2 csc ~/ \10 / x \Ix / CIx / 2 exp ` pl pm r2 csc ~~ Si r m Si r2 exp '- pl pm r2 csc \ m (5) where p/p Im is the mass absorption coefficient for x or Si radiation by the mineral, pm is the mineral density, W is the x-ray take-off angle and (IX /ISi°)r /(Ix°/ISi°)r is the ratio of the generated intensities which is independent of r. The intensity is assumed to be generated at the center of the fiber. Rearranging yields Im ~ a pm cscV(r2-rl) (plm plmi) . 2 259 (6) N 0 a w F+ 0 ~ 0 ~ A
Page 12: bbp52d00 Log in for more options!
This expression provides a satisfactory fit (± 10 percent) to the experimental data in figure 3 except in the case of contamination at small fiber diameters [1]. Equation 6 illustrates that it is the difference between the mass absorption coefficients that deter- mines the magnitude of the absorption effect. When p/p'mineral >>'p/plmineral' a decrease in Iw/ISi occurs with decreasing size because the relative increase in emission will be greater for the element with the larger absorption coefficient. Thus, in grunerite there is a greater relative increase in Si emission (N/pl9~unerite = 1455) than in Fe emission (1'/plgrunerite - 65) and a subsequent 25 percent decrease in I(Fe)/I(Si) as the diameter decreases from 1.5 to 0.15 pm. When N/plmineral << N/p jmineral' Ix/ISi increases with decreasing size because the relative increase in emission is greater for x than for Si. Thus in grunerite, where N/plMg = 3460 and p/pISi = 1455, there is a greater grunerite grunerite relative increase In Mg emission and a subsequent 50 percent increase in I(Mg)/I(SI) as the size decreases from 1.5 to 0.15 pm. The easiest way of correcting for such effects is to use calibration curves of the type shown in figure 3. Combining eqs. (1) and (5) shows that (SA8)til(SA8)tZ = R2/Rl where t is the film thickness (r = t/2). In the case of a very thin film or fiber, taking the limit in eq. (6) as t approaches zero gives: AB not-so-thin , t PB u_ A In A8(thi--~ ~ -pfilm cscy ~ ( plfilm - plfilm ) (7) which is in accord with the expression published recently by Goldstein et al. (23]. The SCu Si' S5n Si and 5Cr Si values used to calculate the Cu-Sn-Cr values were corrected for absorption using SA8 (not-so-thin) values from eq. (7), and in all cases the relative error in concentration decreased as shown in Table 3. Figure 4 can be used as a guide to determine when an absorption correction is advisable. When the absorption coefficient difference for a given particle radius or film thickness is above the line, the absorption correction will be greater than 10 percent and should be taken into account. Many of the amphibole fibers with diameters of 0.2 pm and over require absorption corrections (1]. 260
Page 13: bbp52d00 Log in for more options!
10,000 5000 500 100 0 I 100 200 pfXmtflim/2 in µg/cm2 300 Figure 4. A(pIp) = u/p Ifilmne u/P Ifilmne (pt)film = film mass thickness. When the value of o(u/p) for a particular film thickness is below the line, the absorption correction will be less than 10 percent. The absorption correction will exceed 10 percent for values above the lines. The values shown for amosite and crocidolite indicate that the absorption correction is significant for relatively thin fibers. Instrumentally Induced Contamination Superimposed on the absorption effects just described is the sample contamination which occurs when the hydrocarbons from the vacuum pump fluids are decomposed by the electron beam and deposited on the sample surface [10]. The deposited thickness can, in time, represent an appreciable portion of the total sample thickness. The magnitude of the problem depends upon; 1) the cleanliness of the vacuum system; 2) the electron beam current density; 3) the duration of the analysis; and, 4) the difference in absorption by carbon for the x-ray lines of interest. The magnitude of the latter effect can be estimated from the following expression: with In {Ix/I5i)Woithoutnation = pc tc csc>y pICiK - plC (Ix/ISi)contamination 261 0.2 µm diameter amosite (Mg/Si) **-0.4 pm diameter crocidolite (Fe/Si) I (8)
Page 14: bbp52d00 Log in for more options!
where pC is the density of carbon and tC is the thickness of the carbon deposit in cm. Figure 5 shows the observed variation of IMg/ISi in chrysotile with time for different current densities. The analysis of small (300-400A) chrysotile fibers often requires a small electron beam (higher current density) and a longer analysis time (>5 minutes) to generate credible counting 'statistics. Even though p/pl~~ - N/pI~g is 800, the rapid decrease in IMg/ISi can only be partially accounted for by contamination implying other electron beam induced effects. When the difference in absorption coefficients Is small, contamination is not a serious problem as indicated in figure 5 for the Cu-Cr-Sn film. µ CuKa CrKa SnLa MgKa SiKa p o 5 14 51 1170 360 1.0 018 0.8 0.4 0. I 10 Chrysotlle Fibers 3.7 Km Beam a Chrysotile Fibril 0.6 pm Beam Cu-Sn-Cr Thin Film 0.7 pm Beam i I 20 30 Time In Minutes 1 40 50 Figure 5. Elemental intensity ratios as a function of the duration of electron bombardment in an ATEM operated at 80 keV. IM9JISi and ICr/ISn are plotted for chrysotile asbestos fibers and a Cu-Sn-Cr thin film respectively. The beam diameter for each analysis is indicated on the curves. The mass absorption coefficients for the indicated radiation by carbon are also shown. Optimu. Conditions for Analysis In thin films, theory predicts [24] that the peak-to-background ratio should vary approximately as In U with E , increasing rapidly at low U and then more slowly, where U is the over-voltage ratio, °acceleration potential/excitation potential. This is not always observed experimentally as shown in Table 4. The failure to increase continuously with voltage is, in part, due to the background contribution from extraneous radiation which varies from instrument to instrument. The superiority of the STEM (vs. TEM) configura- tion is indicated in Table 4 where the two STEM instruments have their best peak-to- background ratios at the highest voltage. Unfortunately, fiber or particle counting in the STEM mode is not practical [2]. When column modifications are completed, the optimum operating conditions should be experimentally determined for each instrument. Note that low voltage operation will promote absorption and backscatter effects and reduce the effectiveness of SAEO on thicker fibers. 262
Page 15: bbp52d00 Log in for more options!
Table 4. Experimental determinations of the acceleration potential providing the maximum peak-to-background ratios in the ATEM. E in keV for maximum Investigator Instrument and mode X-ray line opeak to background This report EM 300-TEM CuK 60 This report EM 300-TEM SnL 40 Russ[39] EM 300-TEM FeK 50 Joy & Maher[25] JEOL 100B-STEM M9K 100 Mizuhira[29] JEOL 100C-TEM Na-CiK 20-40 Ga11e et al.[19] Cameca-TEM A1K, Au 20 Geiss & Kyser[27] EM 301-STEM Fe and CuK 100 While there are some mineralogical ambiguities that cannot be resolved by EDS, a well-designed ATEM with the appropriate column modifications used in conjunction with good analytical procedure can provide distinctive mineral spectra that are of great utility in fiber identification. ' Selected Area Electron Diffraction Vastly differing claims have been published as to the utility of SAED in the identifi- cation of mineral fibers: Ampian [28] finds that positive identification using SAED is only forthcoming from carefully indexed patterns yielding accurate lattice parameters. Ross [29] found SAED patterns of asbestos minerals difficult to obtain and interpret and that 200 keV was required to have distinct patterns. Beaman and File [1] reported that only about 10 percent of the chrysotile fibrils examined in a standard gave distinct patterns (40 percent were crystalline). Biles and Emerson [30] reported that most chrysotile fibers in beer did not give identifiable patterns. Samudra [31] reported that 99 percent of the chrysotile fibers in the size range of 200-1200 A provided good patterns. Much of this variation can be accounted for. A distinctive SAED pattern for chrysotile: 1) has a characteristic layer line spacing; 2) is streaked in alternate layer lines; and 3) shows some characteristic reflections, e.g., those in the second row from center are often quite distinctive. We classify as positive only those fibers exhibiting all of these characteristics. Fibers showing only the correct layer line spacing as determined visually on the fluorescent screen are clas- sified as ambiguous; the streaking or characteristic reflections are not sufficiently distinctive to permit positive identification. Patterns without systematic reflections or distinctive layer lines are classified as unknown and the sum of positive, ambiguous, and unknown is termed crystalline. The percentage of fibers in each category has been deter- mined as a function of fiber size using different instruments, standards, and sample preparation methods. Droplets of 10 pL volume, prepared from the dispersion of a high purity chrysotile standard [32] in water, were placed on carbon-coated formvar films on TEM grids. The samples were examined at 00 tilt in a Philips EM300 at 80 keV and a JEOL 100B at 60 and 100 keV. Fiber searching was carried out in the selected area mode with the diffraction aperture in position and focused to minimize the time lapse between finding a fiber and obtaining a SAED pattern. The aperture size at the specimen level was 1-2 pm, the camera length was minimized, and the SAED patterns were focused with the diffraction and objective lens controls. 263 2063105058
Page 16: bbp52d00 Log in for more options!
Figure 6 shows that less than 15 percent of the individual chrysotile fibrils (300- 400 A in diameter) provide positive SAED patterns. A significantly larger portion (20-50 percent) do exhibit the correct layer line spacing (positive + ambiguous) as observed on the fluorescent screen. For the fraction of positive fibers to exceed 50 percent, the fibers must contain over 3 fibrils. 100 80 60 40 20 0 I 2 3 4 5 No. of Chrysotile Fibrils in Fiber e 7 Figure 6. The percentage of chrysotile fibers in a standard providing the indicated quality of the SAEO pattern is shown to depend upon the number of fibriis in the chrysotile fiber. The results obtained on two different instruments are plotted along with previously reported results [1j. All samples were prepared using 10 uLl water droplets containing suspended chrysotile. The results obtained in instrument B were similar at 60 and 100 keV. The lower two curves in figure 6 compare the present results with earlier work [1]. The differences are due to the present use of slightly more stringent requirements for positive identification and possibly to the use of different standards (Wards in reference 1 vs. Union Carbide). Figure 7 illustrates that the percentage of fibers providing diffraction patterns in every category is lower when using samples prepared by the Jaffe extraction of carbon-coated Nuclepore as compared to water droplets. This is presumably due to the carbon coating and/or the presence of some residual Nuclepore. Note that the positive fiber category is not significantly affected by sample preparation. 264
Page 17: bbp52d00 Log in for more options!
as 80 20 0 100 2 Positive + Ambiguous Positive In Water Drop After Jaffe Extraction Of Nuclepore 3 4 5 No. of Chrysotile Fibrils In Fiber a Figure 7. The percentage of chrysotile fibers providing the indicated SAEg pattern quality is shown to depend, to some extent, on the method of sample preparation. The results for 10 pL water droplets are compared with those obtained after Jaffe extraction of a Nuclepore filter in chloroform. All samples were examined in instrument A. 7 The primary reasons for the differing claims are the use of different criterion for classifying a pattern as positive and differences in the fibril content of the fibers being examined. A rigorous definition of positive SAED is needed if identification errors are to be avoided and interlaboratory agreement achieved. Figure 6 shows that over 70 percent of the fibers containing. three fibrils show the correct layer lines spacing (positive + ambiguous category). Most published SAED patterns are not from single fibrils as indicated by the presence of partial rings and diffraction spot smearing or multiplicity [28,33]. To a lesser extent, the reported variation is due to differences in: 1) standard source and treatment; 2) sample preparation methods; 3) instrumental capabilities; 4) operator judgment; and 5) diffraction technique. In the river, tap water, and lake samples we have studied, the chrysotile has consisted predominantly of fibers with 3 or less associated fibrils with single fibrils appearing most frequently. The fibers in 50 percent NaOH produced from chlorine cells using chrysotile asbestos diaphragms are predominantly fibrils and 80 percent have lengths less than 2 pm and 95 percent have lengths less than 5 Ns. Identification based on morphology or SAED alone in these cases has not been particularly reliable because less than 20 percent of the chrysotile fibers had a tubular appearance and only 5-30 percent gave positive SAED patterns. Those fibers identified as chrysotile had EDS spectra and fibril diameters characteristic of chrysotile. 265 2063105060
Page 18: bbp52d00 Log in for more options!
In counting fibers with the ATEM, searching with the diffraction aperture in place is not practical because the field diameter is decreased from about 7 pm to 1 pm. When counting in the TEM mode, the fiber is subjected to more electron beam bombardment before a diffraction pattern can be obtained. When searching with the diffraction aperture in position, the SAED patterns from chrysotile fibers containing three or less fibrils generally fade within 30 seconds to such an extent as to be unidentifiable. This electron beam induced change is due to dehydroxylization [28] and carbon contamination. Reliability of the Method If a sufficient number (typically 60-100) of fibers are analyzed [1,2], the method will generally provide concentrations that are accurate within a factor of two. The reproducibility is considered to be represented by the coefficient of variation or 100o/mean fiber concentration. Inter-laboratory reproducibility between two different Dow laboratories measuring chrysotile in 50 percent NaOH, which is a relatively clean sample, has recently been better than 20 percent (see Table 5). This is reasonably good performance for the small, amount of material being detected as shown in Table 5. The idgntification of an 1000A long chrysotile fibril corresponds to the detection of 3 x 10 18 grams of material [24]. The results will not be this good for a series of labora- tories using a variety of sample preparation techniques and differing criteria for fiber identification. Table 5. Experimentally measured asbestos concentrations. Sample Concentration in millions of fibers per liter Mass of asbestos in parts per billion by weiqht Midland, MI Tap Watera 0.6 0.001 Waste Water Effluenta 10-400 0.2-10 50% NaOHa 50-5000 0.5-40 Duluth Tap Waterb 25 25 50% NaOHa sample 1 Dow Lab A Dow Lab B 380 380 50% NaOH sample 2 380 300 50% NaOH sample 3 530 520 50% NaOH sample 4 1900 1500 a Chrysotile b Amphibole In order to achieve good reproducibility, we adhere to the following: 1. Use a sample preparation method with proven low fiber loss such as the extraction of carbon-coated Nuclepore [2,5,6] or apply a fiber loss correction to each sample [1,2]. 2. Count only samples that have a uniform distribution of solids on the TEM grid, i.e., the fibers per unit area should not fluctuate widely [1,2]. 266
Page 19: bbp52d00 Log in for more options!
C2 3. Count until a sufficient number of fibers (generally 60-100) have been detected so that number of fibers per unit area does not change significantly with additional counting [1,2]. 4. Use a sample volume that provides a particulate density with minimum inters ferences from non-fibrous solids. 5. Modify the TEM column to reduce electron scattering and secondary fluorescence. 6. Subtract off-fiber EDS spectra from fiber spectra. 7. Correct for absorption, when present, using standards or relative sensitivity factors. 8. Minimize contamination rates, when possible, by the use of low current density and short analysis times. 9. Experimentally determine the optimum acceleration potential which often differs for EDS and SAED performance, necessitating a compromise. 10. Use a reasonable and consistent scheme for classifying fibers. The authors wish to thank L. Sturkey and W. A. Knox of The Dow Chemical 6mpany, Walnut Creek, California, for helpful discussions concerning selected area electron dif- fraction and R. H. Geiss and D. F. Kyser of IMB, San Jose, California, for their komments and critical review of the manuscript. The assistance of D. J. Peterson of Dow Chemical of Canada Ltd. and E. B. Bradford of Dow, Midland, in performing the experimental measure- ments is also gratefully acknowledged. References [1] Beaman, D. R. and File, D. M., Anal. Chem. 48, 101 (1976), also in Proceedings Microbeam Analysis Society igth AnnualZ'onference, paper 31 (1975). [2] Beaman, D. R. and Walker,. H. J. , in FDA Symposium on Electron Microscopy of Microfibers (Aug. 1976) in press. [3] Ortiz, L. W. and Isom, B. L. , in 32nd Annual Proceedings of EMSA 554 (1974). [4] Zumwalde, R., In FDA symposium on Electron Microscopy of Microfibers (Aug. 1976) in press. [5] Cook, P. M., Rubin, I. B., Maggiore, C. J., and Nicholson, W. J., in Proceedings of International Conference on Environmental Sensing and Assessment Section 34-1 I.E.E.E. Las Vegas (1976). [6] Anderson, C. H. and Long, J. M., Preliminary Interium Procedure for Fibrous Asbestos, U.S. Environmental Protection Agency, Athens, GA (1977). [7] Benefield, D., The Dow Chemical Company, Freeport, Texas, private communication (1977). [8] Millette, J. R., in FDA Symposium on Electron Microscopy of Microfibers (Aug. 1976) in press. [9] Stewart, I., in FDA Symposium on Electron Microscopy of Microfibers (Aug. 1976) in press. 267 2063105062
Page 20: bbp52d00 Log in for more options!
[10] Beaman, 0. R. and Isasi, J. A., Electron Beam Microanalysis, 5TP506, American Society for Testing and Materials, Philadelp ia 9722T [11] Philibert, J. and Tixier, R., in Physical As ects of Electron Microscopy and Microbeam Analysis; Seigel, B.M. and Beaman, D.R., ds John Wiley and Sons New York, 333 1975). [12] Bolon, R. B. and McConnell, M. D., in Scanning Electron Microscopy/IITR[/SEM/76, Part 1 (1976). [13] Russ, J. C., in Scanning Electron Microscopy/IITRI/SEM/77 1 335 (1977). [14] Joy, 0. C. and Maher, 0. M., in Scanning Electron Microscopy IITRI/SEM/77 1 325 (1977). [15] Zaluzec, N. J. and Fraser, H. L., in Proceedings Microbeam Analysis Society, 11th Annual Conference, paper 14 (1976). (16] Packwood, R. H., Laufer, E. E. , and Roberts, W. N., in Proceedings Microbeam Analysis Society, 12th Annual Conference, paper 115 (1977). [17] Geiss, R. H. and Huang, T. C., X-ray Spectrometry 4 196 (1975). [18] Kyser, D. F. and Geiss, R. H., in Proceedings Microbeam Analysis Society, 12th Annual Conference (1977) paper 110; also private communication with R.H. Geiss (1977). [19] Ouncumb, P., J. de Microscopie 7, 581 (1965). (20] Cliff, G. and Lorimer, G. W., J. Microscopy 103, 203 (1975). [21) Russ, J. C., in Proceedings Microbeam Analysis Society, 8th Annual Conference (1973) paper 30; also in Edax Editor, 5, 11 (1975); also J. Submicr. Cvtol., 6, 55 (1974). [22] Sprys, J. W. and Short, M. A., in Proceedings Microbeam Analysis Society, 11th Annual Conference (1976) paper 9; also private communication with Sprys, J.W. (1977). [23] Goldstein, J. I., Costley, J. L., Lorimer, G. W., and Reed, 5. J. B., in Scanning Electron Microscopy IITRI/SEM/77 1, 315 (1977). [24] Beaman, 0. R., in Modern Techniques for the Detection and Measurement of Environmental Pollutants, 10th Rochester International Conference on Environmental Toxicity (May 1977) in press. [25] Reed, S. J. B. and Ware, N. G., X-ray Spectrometry 3, 149 (1974). [26] Rao, P. and Lifshin, E., in Proceedings Microbeam Analysis Society, 12th Annual Conference, paper 118 (1977). [27] Beaman, D. R. and Solosky, L. F., in Proceedings Microbeam Analysis Society, 9th Annual Conference, paper 26 (1974). [28] Ampian, S. G., in FDA Symposium on Electron Microscopy of Microfibers (Aug. 1976) in press. (29] Ross, M., in FDA Symposium on Electron Microscopy of Microfibers (Aug. 1976) in press. [30] Biles, B. and Emerson, T. R. , Nature, 219, 93 (1968). [31] Samudra, A. V., in Scanning Electron Microscopy IITRI/SEM/77 1(1977). 268
Page 21: bbp52d00 Log in for more options!
SS [32] Union Carbide chrysotile standard provided by K.S. Chopra, Union Carbide Corp., Niagara Falls, NY (1977). [33] Mueller, P. K., Alcocer, A. E., Stanley, R. L., and Smith, G. R., Asbestos Fiber Atlas, Environmental Protection Technology Series, EPA-650/2-75-036 (1975). Discussion K. HEINRICH: When you showed the variation of intensity with fiber diameter, was the scale in micrometers? D. BEAMAN: Yes. P. McGRATH: What can be done to develop criteria to reduce the energy-dispersive interferences so that we can develop criteria for asbestos? BEAMAN: We can do much better with the EDS spectra than in the past by making column modifications and by subtracting background spectra from the fiber spectra. Question (inaudible): BEAMAN: You can make an identification in the STEM mode, but you cannot count fibers easily. It would be difficult to continuously switch from TEM to STEM. C. PARMENTIER: I would like to make a comment concerning TEM-SAED and the lack of d-spaces and difficulty in measuring them for single-fiber chrysotile or amphibole asbestos in small particulates; we run into the same problem of rapidly decreasing signal intensity. We have used a cold finger with liquid nitrogen which allows d-spacings to be resolved on the screen, photographed, and subsequently measured and indexed directly on the negative, so we come up with very accurate d-spacings. The second point I'd like to make is in the spectrometric measurement of Mg-Si ratios. Have you seen varying Mg-Si ratios from chrysotiles of different locals, and is this taken into account in your analysis? BEAMAN: We have used two chrysotile standards, but the chemical differences are smaller than data reproducibility. We could not detect any trend. We, of course, use a cold finger but still observe the rapid deterioration of SAED patterns in the case of chrysotile. Amphibole patterns on the other hand do not tend to fade. N O ~ W 269 0 un ~ .P

Text Control

Highlight Text:

OCR Text Alignment:

Image Control

Image Rotation:

Image Size: